Hydrogen Gas Box Furnace is a large box furnace designed for material synthesis under hydrogen or inert gas atmosphere up to 1700 ºC. The Hydrogen furnace consists of high-quality alumina fiber bricks and three sides heated Mo-alloy coil with a vacuum-sealed steel case with a water cooling jacket. This Gas furnace is an ideal tool for preparing new materials that require a reduced atmosphere。

  • Price
    0
total sales0
evaluate0
Qty
- +

Pledge: genuine guarantee,speed refund,gift insurance,seven days no reason to return

  • Product Details
  • HTVF-1700-A Hydrogen Gas Box Furnace is a large box furnace designed for material synthesis under hydrogen or inert gas atmosphere up to 1700 ºC with a chamber dimension of 400 x 400 x 400 mm. The Hydrogen furnace consists of high-quality alumina fiber bricks and three sides heated Mo-alloy coil with a vacuum-sealed steel case with a water cooling jacket and automatic H2 burning system. This Gas furnace is an ideal tool for preparing new materials that require a reduced atmosphere, such as phosphorous and Ti alloy, and also heat-treating materials under Ar gas.

     

    Furnace Structure

    Double-layer stainless steel structure for maximum vacuum and safety, which can ensure a positive pressure up to 0.06MP and vacuum up to -0.1Mpa

    Water cooling front door and air cooling shell

    Built-in pressure gauge, vacuum valves, and two gas float meters

    Built-in H2 burning control system --- if the controller detects no flame in the burner, the gas delivery valve will be shut down automatically    

    Inside Chamber 

    Heated chamber: 400 x 400 x 400 mm, 65 Liter (Max)

    Energy-saving 1800 grade fibrous alumina insulation

    Sealed by stainless steel with water cooling

    Three sides heating (Left, right and back) to ensure the best temperature uniformity.

    SiC plate is installed on the bottom, which has 100kg, Max. loading

    Heating Elements

    9 sets of Mo coils heating elements are built-in and replaceable 

    The Mo heating element is consumable. You may order spare element by click Pic. 1 below

    Warning:  Mo heating element can not be used under air or oxygen at the temperature > 300°C

    The heating element can be replaced by 1900 grade MsSi2 to run the furnace under oxygen and air and Inert gases up to 1700°C ,

    Power

    36 KW

    Voltage

    AC 208 - 240V three phases (50/60 Hz)

    Current

    80 Amp (100 Amp air breaker is required)

    Continuous Working Temperature

    1650° C    ( in H2 gas or Ar gas, not recommended under N2 gas )

    Maximum Working Temperature

    1700°C (< 1 hr)   ( in H2 gas or Ar gas, not recommended under N2 gas )

    Heating Rate

    5°C / min

    Temperature Accuracy

    +/- 0.1°C

    Temperature Control

    and Processing control

    Temperature and processing control via touch screen computer, including

    Temperature control with +/- 1ºC accuracy.

    Two thermal couples are built-in: one for temperature control and one for safety monitoring 

    30 programmable segments for precise control of heating rate, cooling rate, and dwell time.

    Over-temperature protection by an independent controller

    Recording and display temperature profile during operation

    Processing control includes

    Purging procedure for the furnace chamber

    Chamber pressure control

    Gas flow control

    Recording and display pressure profile during operation

    Gas Flow Control

    and Vacuum Valve

    Gas Flow Control and Vacuum ValveTwo large flow meters (N2: 0~ 12L/min, H2: 0~12L/min) are installed in the front panel of the furnace.

    One for purging Inert gas prior to introducing H2 gas

    One for controlling H2 gas flow

    KF40 vacuum pumps built-in to connect a vacuum pump

    The vacuum pump may be needed, but not included

    H2 Gas Burning Control System

    An ignition device already installed to incinerate flowing gas and detect burner temperature to ensure H2 burner is in working condition. If H2 does not burn properly, the controller will shut down the H2 gas delivery valve immediately.

    Warning:  You must install a certified Hydrogen detector in operation facility to alarm hydrogen leaking b accident. Please click the link below to order a hydrogen detector from us at extra cost.

    Water Cooling

    A 16L/min recirculating water chiller is installed a bottom furnace for cooling furnace sealing flange. 

    Overall Dimensions

    1600X1320X2100 mm (not including transformer and water chiller)

    680X550X730 mm  ( transformer )

    Shipping Weight

    1000 kg

    Warranty

    One-year limited warranty with lifetime support.  (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.)

    ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of One Year Limited Warranty.

    Attention: A free course of safety operation training at the MTI field is required to purchase this furnace. A customer's side training is also available upon request at extra cost. If the users believe that they have sufficient experience and knowledge to handle the furnace in the dangerous or emergent situation, a liability release agreement must be signed at the purchase. HT Corp. is not responsible for any damages caused by misuse without safety operation training. Any damages caused by the corrosive and acidic gases are not under the coverage of the One Year Limited Warranty.

    Application Notes

    The pressure inside the chamber should ALWAYS be controlled at 0.01-0.02 Mpa above the standard atmospheric pressure during the entire heating process, otherwise, the oxygen in the air may backflow into the furnace chamber and cause an explosion in the worst-case scenario.

    ALWAYS place the furnace in a room with the fume hood and a hydrogen detector installed.

    Good ventilation is necessary for the room in which place the furnace

    Exhausting Hydrogen gas in the room is forbidden.

     


    HTVF-1700-A Hydrogen Gas Box Furnace is a large box furnace designed for material synthesis under hydrogen or inert gas atmosphere up to 1700 ºC with a chamber dimension of 400 x 400 x 400 mm. The Hydrogen furnace consists of high-quality alumina fiber bricks and three sides heated Mo-alloy coil with a vacuum-sealed steel case with a water cooling jacket and automatic H2 burning system. This Gas furnace is an ideal tool for preparing new materials that require a reduced atmosphere, such as phosphorous and Ti alloy, and also heat-treating materials under Ar gas.

     

    Furnace Structure

    Double-layer stainless steel structure for maximum vacuum and safety, which can ensure a positive pressure up to 0.06MP and vacuum up to -0.1Mpa

    Water cooling front door and air cooling shell

    Built-in pressure gauge, vacuum valves, and two gas float meters

    Built-in H2 burning control system --- if the controller detects no flame in the burner, the gas delivery valve will be shut down automatically    

    Inside Chamber 

    Heated chamber: 400 x 400 x 400 mm, 65 Liter (Max)

    Energy-saving 1800 grade fibrous alumina insulation

    Sealed by stainless steel with water cooling

    Three sides heating (Left, right and back) to ensure the best temperature uniformity.

    SiC plate is installed on the bottom, which has 100kg, Max. loading

    Heating Elements

    9 sets of Mo coils heating elements are built-in and replaceable 

    The Mo heating element is consumable. You may order spare element by click Pic. 1 below

    Warning:  Mo heating element can not be used under air or oxygen at the temperature > 300°C

    The heating element can be replaced by 1900 grade MsSi2 to run the furnace under oxygen and air and Inert gases up to 1700°C ,

    Power

    36 KW

    Voltage

    AC 208 - 240V three phases (50/60 Hz)

    Current

    80 Amp (100 Amp air breaker is required)

    Continuous Working Temperature

    1650° C    ( in H2 gas or Ar gas, not recommended under N2 gas )

    Maximum Working Temperature

    1700°C (< 1 hr)   ( in H2 gas or Ar gas, not recommended under N2 gas )

    Heating Rate

    5°C / min

    Temperature Accuracy

    +/- 0.1°C

    Temperature Control

    and Processing control

    Temperature and processing control via touch screen computer, including

    Temperature control with +/- 1ºC accuracy.

    Two thermal couples are built-in: one for temperature control and one for safety monitoring 

    30 programmable segments for precise control of heating rate, cooling rate, and dwell time.

    Over-temperature protection by an independent controller

    Recording and display temperature profile during operation

    Processing control includes

    Purging procedure for the furnace chamber

    Chamber pressure control

    Gas flow control

    Recording and display pressure profile during operation

    Gas Flow Control

    and Vacuum Valve

    Gas Flow Control and Vacuum ValveTwo large flow meters (N2: 0~ 12L/min, H2: 0~12L/min) are installed in the front panel of the furnace.

    One for purging Inert gas prior to introducing H2 gas

    One for controlling H2 gas flow

    KF40 vacuum pumps built-in to connect a vacuum pump

    The vacuum pump may be needed, but not included

    H2 Gas Burning Control System

    An ignition device already installed to incinerate flowing gas and detect burner temperature to ensure H2 burner is in working condition. If H2 does not burn properly, the controller will shut down the H2 gas delivery valve immediately.

    Warning:  You must install a certified Hydrogen detector in operation facility to alarm hydrogen leaking b accident. Please click the link below to order a hydrogen detector from us at extra cost.

    Water Cooling

    A 16L/min recirculating water chiller is installed a bottom furnace for cooling furnace sealing flange. 

    Overall Dimensions

    1600X1320X2100 mm (not including transformer and water chiller)

    680X550X730 mm  ( transformer )

    Shipping Weight

    1000 kg

    Warranty

    One-year limited warranty with lifetime support.  (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.)

    ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of One Year Limited Warranty.

    Attention: A free course of safety operation training at the MTI field is required to purchase this furnace. A customer's side training is also available upon request at extra cost. If the users believe that they have sufficient experience and knowledge to handle the furnace in the dangerous or emergent situation, a liability release agreement must be signed at the purchase. HT Corp. is not responsible for any damages caused by misuse without safety operation training. Any damages caused by the corrosive and acidic gases are not under the coverage of the One Year Limited Warranty.

    Application Notes

    The pressure inside the chamber should ALWAYS be controlled at 0.01-0.02 Mpa above the standard atmospheric pressure during the entire heating process, otherwise, the oxygen in the air may backflow into the furnace chamber and cause an explosion in the worst-case scenario.

    ALWAYS place the furnace in a room with the fume hood and a hydrogen detector installed.

    Good ventilation is necessary for the room in which place the furnace

    Exhausting Hydrogen gas in the room is forbidden.